Barco MCD214ACU Monitor

Barco MCD214ACU is a medical-grade CRT monitor used in ultrasound imaging systems. Common failure modes include loss of high‑voltage output due to failed flyback transformer, HV tripler assembly, or horizontal output xistor; these components often exhibit pop sounds and brief initial display before blanking. First‑check diagnostics: verify mains power and 12 V DC rail continuity, inspect the flyback winding for shorts or opens, test the HV tripler capacitors/diodes with a multimeter, and confirm the horizontal output xistor’s voltage rating and leakage current.[oss-gpt-20b]

Need your Barco repaired or tested?

We specialize in the repair and refurbishment of high-end industrial control electronics and power systems like this unit.

📞 Call Now: 1 (336) 969-0110

Free Estimates • Industrial-grade testing • Nationwide Service

Call to Purchase AI Services — help us build the next Grace Blackwell GX10 node.

Industrial AI Hardware Recovery — Free Diagnostic • Pay Return Shipping Only

[oss-gpt-20b]

We reclaim and repair GPUs and industrial AI hardware (cards, PSUs, fan modules, VRM boards). Free estimate — you only pay return shipping if you decline.

📞 Call: 1 (336) 969-0110 ✍️ Start Free Diagnostic

Free Estimates • Industrial-grade testing • 1-Year Warranty • Nationwide Service


Other Repaired Units by Barco

Barco MCD214ACU Monitor
Barco DP-30 power supply
Barco DP-30 Power Supply
Barco R763684-00 Circuit Board
Barco R76-22935 power supply

Contact Us

🚑Shop Location🛠️

Industrial Electronic Repair
145 Rockwell Place Ct
Tobaccoville, NC 27050
🚑 Rush repair shipments only — no walk-ins.

📞Phone

1-336-969-0110

📦Mailing Address🚛

Industrial Electronic Repair
931-E S. Main St #204
Kernersville, NC 27284
📦 Mailing Address (for invoices or shipped repairs)
No walk-ins accepted.