Barco MCD214ACU is a medical-grade CRT monitor used in ultrasound imaging systems. Common failure modes include loss of high‑voltage output due to failed flyback transformer, HV tripler assembly, or horizontal output xistor; these components often exhibit pop sounds and brief initial display before blanking. First‑check diagnostics: verify mains power and 12 V DC rail continuity, inspect the flyback winding for shorts or opens, test the HV tripler capacitors/diodes with a multimeter, and confirm the horizontal output xistor’s voltage rating and leakage current.[oss-gpt-20b] |